IMP,TSV,3.7,11.5,MTX,MG
Report
- Report Number
- 0002023141-2024-02875
- Event Type
- Injury
- Date Received
- September 9, 2024
- Date of Event
- August 21, 2024
- Report Date
- January 21, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019935
- PMA / PMN Number
- K101977
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM D4: ADDITIONAL DEVICE INFORMATION D9: DEVICE AVAILABILITY G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H4: DEVICE MANUFACTURER DATE H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1250495. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1250495, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : MEDICAL : BONE LOSS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. H3 OTHER TEXT: SUMMARY INVESTIGATION.
ZIMVIE COMPLAINT NUMBER (B)(4).
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 29 WAS REMOVED DUE TO BONE LOSS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1536885 | IMP,TSV,3.7,11.5,MTX,MG | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1250495 | 00889024019935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female | Required Intervention |