FDA Adverse Event
Malfunction
Summary report: N
HOOK CEV2295A 3PK N5 FOR HOOK HANDLE
MDR report key: 20170427
·
Received September 9, 2024
Report
- Report Number
- 3003249645-2024-00044
- Event Type
- Malfunction
- Date Received
- September 9, 2024
- Date of Event
- July 24, 2024
- Report Date
- September 9, 2024
- Manufacturer
- INTEGRA MICROFRANCE S.A.S.
- Product Code
- GEI
- UDI-DI
- 10381780092551
- PMA / PMN Number
- K210942
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE 3PK N5 HOOK FOR HOOK HANDLE (CEV2295A) WAS RETURNED FOR EVALUATION. FAILURE ANALYSIS: EVALUATION OF THE HOOKS VERIFIED THE COMPLAINT REPORTED BY THE CUSTOMER AS VALID. THE COATING OF THE COELIO HOOKS WERE MELTED, ESPECIALLY ON THE TIP SIDE. ROOT CAUSE ANALYSIS: IT WAS DETERMINED THAT THE REASON FOR THE REPORTED FAILURE WAS LIKELY DUE TO THE APPLICATION OF A VOLTAGE THAT WAS TOO HIGH.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE COELIO HOOKS/3PK N5 HOOK FOR HOOK HANDLE (CEV2295A) MELTED DURING USE. NO INJURY OR SURGICAL DELAY HAS BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1510410 | HOOK CEV2295A 3PK N5 FOR HOOK HANDLE | PFM16 | GEI | INTEGRA MICROFRANCE S.A.S. | 10381780092551 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |