FDA Adverse Event Injury Summary report: N

NOA MEDICAL TALL SOFT TOUCH SIDE RAIL

MDR report key: 2017022 · Received March 2, 2011

Report

Report Number
1931947-2011-00001
Event Type
Injury
Date Received
March 2, 2011
Date of Event
January 27, 2011
Report Date
February 17, 2011
Manufacturer
NOA MEDICAL INDUSTRIES, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

NO ACTUAL INJURIES OCCURRED BUT THE POSSIBILITY OF INJURY WAS PRESENT. THE THREADS ON 1 OF 2 THREADED ATTACHMENT PINS WERE STRIPPED. WHEN THE 550 LBS RESIDENT ROLLED AGAINST THE SIDE RAIL, THE STRIPPED THREAD LET GO, OR PULLED OUT, CAUSING PIN 2 OF 2 TO FAIL (SNAP OFF). NOA HAS IMPROVED THE DESIGN OF THE "TALL SOFTTOUCH SIDE RAIL" P.N 7540011BEI TO ELIMINATE THE NEED FOR THE CUSTOMER TO INSTALL AND PERFORM PERIODIC INSPECTION AND MAINTENANCE ON THE THREADED ATTACHMENT PINS. THE NEW DESIGN SIDE RAIL, P.N 7540027BEI HAS FACTORY INSTALLED, WELDED, MAINTENANCE FREE ATTACHMENT PINS. THE NEW DESIGN IS CURRENTLY SHIPPING FOR ALL BEDS RATED OVER 600 LBS.

Description of Event or Problem · 1

LONG TERM CARE RESIDENT APPROX 550 LBS TURNED TO LEFT SIDE FOR BATHING AND RAIL "SNAPPED". RESIDENT ON FLOOR. ADMITTED TO HOSPITAL. NO INJURIES WERE FOUND. SHE WAS RELEASED FROM THE HOSPITAL THE NEXT MORNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOA MEDICAL TALL SOFT TOUCH SIDE RAIL HOSPITAL BED SIDE RAIL FNL NOA MEDICAL INDUSTRIES, INC. 7540011BEI NM21

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization