FDA Adverse Event Malfunction Summary report: N

TSRH 3DX

MDR report key: 20169700 · Received September 9, 2024

Report

Report Number
1030489-2024-01104
Event Type
Malfunction
Date Received
September 9, 2024
Date of Event
August 12, 2024
Report Date
November 10, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PART# 8350392 LOT# EM09K002 VISUAL AND OPTICAL EXAMINATION IDENTIFIED IT APPEARS THAT PART OF THE TIP OF THE INSTRUMENT HAS BEEN BROKEN OFF. THE METAL DEFORMATION GIVES INDICATION THAT THE FAILURE WAS THE RESULT OF OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM USER FACILITY (UF) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT WILL BE USED IN SPINAL THERAPY. IT WAS REPORTED THAT THE TIP OF THE DRIVER WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT AND NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1536618 TSRH 3DX ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH MEDTRONIC SOFAMOR DANEK USA, INC 8350392 EM09K002

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown