FDA Adverse Event Malfunction Summary report: N

MEDTRONIC NAVIGATION, INC

MDR report key: 20159201 · Received September 6, 2024

Report

Report Number
1723170-2024-02566
Event Type
Malfunction
Date Received
September 6, 2024
Date of Event
August 20, 2024
Report Date
November 20, 2024
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00613994480637
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3, H6) THE INSTRUMENT RETURNED FOR ANALYSIS. THE REPORTED ISSUE WAS CONFIRMED. WHEN CONNECTED TO A KNOWN GOOD SYSTEM IT WAS FOUND THAT LIGHT EMITTING DIODE (LED) #3 WAS NOT FIRING. OTHERWISE, THE RETURNED FRAME DISPLAYED GOOD GEOMETRY AND DIVOT ERROR READINGS WITH NORMAL VERIFICATION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3) NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. CODES: B17, C20, D15 G2) FOREIGN COUNTRY: JAPAN MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT ONLY ONE SENSOR DID NOT RESPOND, IT WAS STILL IN RED.  THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1864873 MEDTRONIC NAVIGATION, INC NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9733881 140812 00613994480637

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown