ANGLED REAMER, DRIVE SHAFT
Report
- Report Number
- 1220246-2024-07430
- Event Type
- Malfunction
- Date Received
- September 6, 2024
- Date of Event
- August 13, 2024
- Report Date
- June 24, 2025
- Manufacturer
- ARTHREX, INC.
- Product Code
- LXH
- UDI-DI
- 00888867324954
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- 501
Narratives
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
ADDITIONAL INFORMATION: D9, G3, H3, H6. COMPLAINT ALLEGATION IS CONFIRMED. ONE UNPACKAGED, AR-9676, BATCH 022338, WAS RECEIVED FOR INVESTIGATION. VISUAL INSPECTION NOTED SIGNS OF WEAR ON THE DEVICE: CIRCUMFERENTIAL GRINDING ON THE DISTAL AND PROXIMAL TIP. FUNCTIONAL TESTING WAS PERFORMED WITH A MATING PART AND IT WAS FOUND THAT THE REAMER CANNOT BE MOVED FREELY. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO MISUSE DUE TO INTERFERENCE WITH THE MATING INSTRUMENT.
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9676 ANGLED REAMER, DRIVE SHAFT, AND AN AR-9597-10 ANGLED REAMER SLEEVE HAD AN ISSUE WHEN REAMING THE GLENOID. WHILE REAMING, THE SURGEON FELT THE REAMER "LOCK UP" ON HIM. THE SURGEON RETURNED THE INSTRUMENTATION TO THE TECH TO INSPECT, AND THE INNER SLEEVE WAS STUCK WITHIN THE OUTER SLEEVE OF THE ANGLED REAMER SHAFT. THIS ISSUE CAUSED THE ENTIRE DEVICE TO ROTATE AS ONE WHEN ON POWER AND PREVENTED THE REAMER FROM WORKING PROPERLY. PLIERS WERE USED TO DISASSOCIATE THE TWO PIECES FROM EACH OTHER. THEY REASSEMBLED IT AND MADE IT WORK TO PROCEED WITH THE CASE. DUE TO THE MALFUNCTION, THIS ISSUE ADDED 15-20 MINUTES TO THE CASE TIME. THIS WAS DISCOVERED DURING AN RTSA PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2140563 | ANGLED REAMER, DRIVE SHAFT | MANUAL INSTR, GENERAL SURGICAL | LXH | ARTHREX, INC. | ANGLED REAMER, DRIVE SHAFT | 022338 | 00888867324954 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |