INTERA 1.5T MASTER/NOVA

MDR report key: 2015790 · Received March 9, 2011

Report

Report Number
3003768277-2011-00266
Event Type
Other
Date Received
March 9, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
LNH
PMA / PMN Number
K001796
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - THE INCIDENT REPORTED IS MOST LIKELY CAUSED BY SKIN-SKIN CONTACT BETWEEN THE INNER THIGHS. THE FACT THAT SEVERAL SANS WITH HIGH SARS VALUES WERE USED MAY HAVE CONTRIBUTED AS WELL. SKIN-SKIN BURNS ARE A KNOWN CAUSE FOR RF BURNS DURING AN MRI SCAN. THE BEST WAY TO PREVENT SKIN-SKIN RF BURNS IS TO CREATE ENOUGH ISOLATION BETWEEN THE TWO SKIN SURFACES EITHER BY DISTANCE OR BY AN ISOLATING MATERIAL BETWEEN THE SKINS. THEREFORE AN ACCESSORY SET IS PART OF EVERY MR SYSTEM DELIVERY CONTAINING SEVERAL SPACERS AND CUSHIONS. THE INSTRUCTIONS FOR USE CONTAIN EXTENSIVE WARNINGS AND INSTRUCTIONS FOR PT POSITIONING.

Description of Event or Problem · 1

THE PT WAS SCANNED FOR A PELVIS EXAMINATION WITH THE QBC POSITIONED HEAD FIRST, SUPINE. ALSO THE SPINE COIL WAS POSITIONED FOR A LUMBAR STUDY, HOWEVER WAS NOT IN USE AND NOT CONNECTED AT THE TIME OF THE BURN. IMMEDIATELY AFTER THE EXAMINATION 2 SECOND DEGREE BURNS OVER 1CM IN DIAMETER ON BOTH THE INNER THIGHS WERE OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERA 1.5T MASTER/NOVA LNH PHILIPS HEALTHCARE 781106

Patients

Seq Age Sex Outcome Treatment
1 77 YR Other