FDA Adverse Event Injury Summary report: N

BALL HEADS: COCR BALL HEAD 12/14 Ø 28 SIZE M 0

MDR report key: 20157866 · Received September 6, 2024

Report

Report Number
3005180920-2024-00732
Event Type
Injury
Date Received
September 6, 2024
Date of Event
August 14, 2024
Report Date
September 6, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805110
PMA / PMN Number
K072857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 23 AUG 2024 LOT 2307377: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10-MAY-2023. EXPIRATION DATE: 2028-04-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 52 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL COMPONENTS INVOLVED: BATCH REVIEW PERFORMED ON 23 AUG 2024. BALL HEADS: BIPOLAR HEAD 25060.2851 BIPOLAR HEAD Ø28X51 (K091967) LOT 2245354: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-MAY-2023. EXPIRATION DATE: 2028-04-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BATCH REVIEW PERFORMED ON 27 AUG 2024 STEM: MASTERLOC 01.39.411 CEMENTLESS TI COATED LAT PLUS STEM SIZE 11 (K173267) LOT 2304304: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-JUL-2023. EXPIRATION DATE: 2028-06-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT 2 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE MEDACTA COCR, BIPOLAR HEAD AND STEM AND REVISED THE COMPETITOR CUP AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2041681 BALL HEADS: COCR BALL HEAD 12/14 Ø 28 SIZE M 0 FEMORAL HEAD JDI MEDACTA INTERNATIONAL SA 2307377 07630030805110

Patients

Seq Age Sex Outcome Treatment
1 87 YR Male Required Intervention