FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 20151216 · Received September 5, 2024

Report

Report Number
1645337-2024-10323
Event Type
Injury
Date Received
September 5, 2024
Date of Event
June 20, 2024
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000327
PMA / PMN Number
P030053
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON SEPTEMBER 11, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE CORRECT DATE OF EXPLANTATION WAS ALSO PROVIDED AS (B)(6) 2024. ON SEPTEMBER 18, 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING THE VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SM MPP GEL 400CC RETURNED DEVICE. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY'S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULES, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION, NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Additional Manufacturer Narrative · 0

AN ANALYSIS OF THE SUSPECT MEDICAL DEVICE¿S PHOTO WAS COMPLETED. INVESTIGATION SUMMARY: UPON VISUAL EVALUATION OF THE IMAGE PROVIDED IN THE COMPLAINT NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION, NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 400CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED BREAST CAPSULAR CONTRACTURES (BAKER GRADE III). AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. THE REPLACEMENT DEVICES WERE: (LEFT) 375CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3503751BC LOT: 9990154 SN: (B)(6) AND (RIGHT) 375CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3503751BC LOT: 9990154 SN: (B)(6). THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2141135 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 9818542 00081317000327

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female Required Intervention