EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 13MM LENGTH
Report
- Report Number
- 0002023141-2024-02856
- Event Type
- Injury
- Date Received
- September 5, 2024
- Date of Event
- June 1, 2023
- Report Date
- January 22, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024343894
- PMA / PMN Number
- K142082
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TC
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1220689. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1220689 FOR SIMILAR EVENTS AND 1 OTHER RELEVANT COMPLAINT(S) WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : MEDICAL : BONE LOSS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #UL2 WAS PLACED WITH IMMEDIATE LOADING, BUT THE CLINICIAN THINKS EXCESSIVE FORCE CAUSED BONE LOSS AROUND THE IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1836542 | EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 13MM LENGTH | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1220689 | 00889024343894 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Female | Required Intervention |