FDA Adverse Event Malfunction Summary report: N

INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE

MDR report key: 20146654 · Received September 5, 2024

Report

Report Number
9611594-2024-00181
Event Type
Malfunction
Date Received
September 5, 2024
Date of Event
August 7, 2024
Report Date
October 9, 2025
Manufacturer
AVANOS MEDICAL INC.
Product Code
KGC
PMA / PMN Number
K080253
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SAMPLE IS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED BY THE MANUFACTURER. A REVIEW OF THE DEVICE HISTORY RECORD AND UDI ARE IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 04 SEP 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT COMP (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE REPORT HAS BEEN RETURNED AND THE INVESTIGATION REMAINS IN PROGRESS AT THIS TIME. ALL INFORMATION REASONABLY KNOWN AS OF 24 SEP 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9, H6. THE PRODUCT INVOLVED IN THE REPORT HAS BEEN RETURNED AND THE INVESTIGATION REMAINS IN PROGRESS AT THIS TIME. ALL INFORMATION REASONABLY KNOWN AS OF 30 SEP 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR THE REPORTED LOT NUMBER, 30273687, IN THIS COMPLAINT WAS REVIEWED AND THE MATERIAL WAS PRODUCED ACCORDING TO THE MANUFACTURING PROCEDURES AND MET THE QUALITY REQUIREMENTS. THE SUTURE WAS RECEIVED WITH THE SOFTSHELL INTACT INTO THE CLAMP BASE. NO T-BAR ASSEMBLY, NO OVERHAND KNOTS ON BOTH ENDS OF SUTURE. WHEN MEASURING, THERE WAS APPROXIMATELY 8MM OF ABSORBABLE SUTURE LEFT ON THE SOFTSHELL BASE SIDE OF THE COMPONENT AND 5MM OF ABSORBABLE SUTURE LEFT ON THE LOCKING CLIP SIDE OF THE COMPONENT. IT WAS APPARENT THAT THE SUTURE BROKE OFF OF THE ANCHOR T-BAR ASSEMBLY, BASED ON THE SUTURES LEFT. ROOT CAUSE COULD NOT BE DETERMINED. ALL INFORMATION REASONABLY KNOWN AS OF 08-OCT-2025 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Description of Event or Problem · 0

AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF THREE REPORTS. REFER TO 9611594-2024-00182 FOR THE SECOND REPORT, REFER TO 9611594-2024-00183 FOR THE THIRD REPORT. IT WAS REPORTED, THE T-FASTENERS ARE USED TO HOLD THE STOMACH TO THE WALL IN ORDER TO PLACE THE GASTROSTOMY. THE THREE T-FASTENERS BROKE WHEN THE INTRODUCER WAS INSERTED. PEDIATRIC DIGESTIVE SURGEONS HAD TO INTERVENE TO APPLY STITCHES TO PREVENT THE RISK OF PNEUMOPERITONEUM DUE TO THE LOOSENING OF THE T-FASTENERS. ACTIONS TAKEN IN THE CARE FACILITY FOR THE PATIENT'S CARE WERE, ¿CALL THE PEDIATRIC SURGEONS AND OPEN ANOTHER INTRODUCTION KIT TO REPLACE OTHER PLOTS [T-FASTENERS].¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2073511 INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE DH EF PERC PLACEMENT PRODUCTS KGC AVANOS MEDICAL INC. 98431 30273687

Patients

Seq Age Sex Outcome Treatment
1 6 YR Female