AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-24197
- Event Type
- Malfunction
- Date Received
- September 5, 2024
- Date of Event
- July 28, 2024
- Report Date
- August 9, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4) SINCE NO LOT NUMBER IS AVAILABLE, A DETAILED INVESTIGATION, TESTS OR REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONCLUDED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH PMS PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WILL FLAG ON THE TRIPS AND ALERTS ACCORDING TO THE OMQR PROCEDURE.
REFERENCE NUMBER 1963822 EVENT OCCURRED IN THE UNITED STATES . ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT ENCOUNTERED TWO INFUSION SETS CANNULA KINKED EVENTS WITHIN 3 HOURS AFTER INSERTION. THE SITE OF INSERTION WAS ABDOMEN.. THE INFUSION SET WAS IN USE FOR 6HRS. DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1836224 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001680 | UNKNOWN | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Male |