FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 2014295 · Received March 10, 2011

Report

Report Number
2017865-2011-01458
Event Type
Injury
Date Received
March 10, 2011
Date of Event
September 21, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER REPLACING THE PATIENT'S IMPLANT- ABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND SUTURING THE POCKET, SENSING AND THRESHOLD MEASUREMENTS -DID NOT MAKE SENSE-. UPON OPENING THE POCKET, THE LEADS WERE FOUND TO BE REVERSED IN THE ICD'S CONNECTOR PORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention (B)(4)