FDA Adverse Event Malfunction Summary report: N

ISOFLEX S

MDR report key: 2014157 · Received March 10, 2011

Report

Report Number
2017865-2011-01425
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
January 4, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT VENTRICULAR R-WAVES DECREASED FROM 12 MV TO 2-3 MV. VENTRICULAR CAPTURE WAS LIMITED AT A MAXIMUM OUTPUT OF 7.5 V, 1.5 MS. X-RAY INDICATED THAT THE LEAD INSULATION WAS POSSIBLY ABRADED. SUBCLAVIAN CRUSH WAS ALSO SUSPECTED. THE LEAD WILL BE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISOFLEX S PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1646T/58 NA

Patients

Seq Age Sex Outcome Treatment
1