FDA Adverse Event Injury Summary report: N

ISOFLEX S

MDR report key: 2014156 · Received March 10, 2011

Report

Report Number
2017865-2011-01427
Event Type
Injury
Date Received
March 10, 2011
Date of Event
December 20, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED FINAL ANALYSIS FOUND THAT THE SUTURE SLEEVE AND THE PROXIMAL INSULATION/COIL WERE DAMAGED DUE TO EXCESSIVE FORCE BEING USED WHEN TYING THE SUTURE SLEEVE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN PERFORMING SURGERY DUE TO PULSE GENERATOR MALFUNCTION, IT WAS NOTED THAT THE LEAD'S PROXIMAL INSULATION WAS DAMAGED/ABRADED. THE LEAD WAS REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISOFLEX S PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1646T/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention