FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 2013997 · Received March 10, 2011

Report

Report Number
2017865-2011-01577
Event Type
Injury
Date Received
March 10, 2011
Date of Event
November 19, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FINAL ANALYSIS FOUND THAT THE DISTAL INSULATION WAS DAMAGED AT 11.5 CM FROM THE CONNECTOR PIN. THE DAMAGED INSULATION CREATED A SHORT BETWEEN THE COILS AND RESULTED IN THE REPORTED IMPEDANCE ANOMALY. THE DAMAGE IS CONSISTENT WITH MECHANICAL STRESS (I.E.: SUTURE SLEEVE TIE-DOWN).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VENTRICULAR LEAD EXHIBITED LOW LEAD IMPEDANCE AND NOISE. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention