FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 2013997
·
Received March 10, 2011
Report
- Report Number
- 2017865-2011-01577
- Event Type
- Injury
- Date Received
- March 10, 2011
- Date of Event
- November 19, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FINAL ANALYSIS FOUND THAT THE DISTAL INSULATION WAS DAMAGED AT 11.5 CM FROM THE CONNECTOR PIN. THE DAMAGED INSULATION CREATED A SHORT BETWEEN THE COILS AND RESULTED IN THE REPORTED IMPEDANCE ANOMALY. THE DAMAGE IS CONSISTENT WITH MECHANICAL STRESS (I.E.: SUTURE SLEEVE TIE-DOWN).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE VENTRICULAR LEAD EXHIBITED LOW LEAD IMPEDANCE AND NOISE. THE LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Required Intervention |