FDA Adverse Event Malfunction Summary report: N

ISOFLEX S

MDR report key: 2013983 · Received March 10, 2011

Report

Report Number
2017865-2011-01404
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
December 16, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VENTRICULAR LEAD EXHIBITED NOISE. EGMS REVEALED NOISE ON BOTH THE ATRIAL AND VENTRICULAR CHANNELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISOFLEX S PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1642T/52 NA

Patients

Seq Age Sex Outcome Treatment
1 (B)(4)