FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 2013887 · Received March 10, 2011

Report

Report Number
2017865-2011-01534
Event Type
Injury
Date Received
March 10, 2011
Date of Event
December 15, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED LOSS OF SENSING. A CHEST X-RAY FOUND THAT THE LEAD HAD DISLODGED INTO THE SUPERIOR VENA CAVA. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention (B)(4)