MERIT CUSTOM KIT
Report
- Report Number
- 1721504-2011-00045
- Event Type
- Malfunction
- Date Received
- February 4, 2011
- Date of Event
- December 8, 2010
- Report Date
- January 13, 2011
- Manufacturer
- MERIT MEDICAL SYSTEMS, INC.
- Product Code
- DQO
- PMA / PMN Number
- K913682
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION: ONE USED SUSPECT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXAMINED VISUALLY AND THE COMPLAINT WAS CONFIRMED. THE ROTATOR WAS SEPARATED AT THE BOND BETWEEN THE COLLAR AND THE HOUSING. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO EXCEPTION DOCUMENTS. THE COMPLAINT DATA BASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE ROOT CAUSES OF THE FAILURE ARE ATTRIBUTED TO THE MANUFACTURING BONDING PROCESS. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THIS TYPE OF FAILURE. THIS SUBASSEMBLY WAS BUILT PRIOR TO IMPLEMENTATION OF THE NOTED CORRECTIVE ACTIONS. COMPLAINT DATA WILL BE MONITORED TO VERIFY EFFECTIVENESS OF CORRECTIVE ACTIONS TAKEN. EVALUATION: DEVICE HISTORY RECORD WAS REVIEWED, COMPLAINT DATA BASE WAS REVIEWED.
THE ROTATOR ON THE STOPCOCK BROKE DURING A TRANSCATHETER HEPATIC ARTERY CHEMOEMBOLIZATION. NO HARM OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MERIT CUSTOM KIT | CATHETER, INTRAVASCULAR, DIAGNOSTIC | DQO | MERIT MEDICAL SYSTEMS, INC. | F756931 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONTRAST MEDIA |