FDA Adverse Event Malfunction Summary report: N

MERIT CUSTOM KIT

MDR report key: 2013815 · Received February 4, 2011

Report

Report Number
1721504-2011-00045
Event Type
Malfunction
Date Received
February 4, 2011
Date of Event
December 8, 2010
Report Date
January 13, 2011
Manufacturer
MERIT MEDICAL SYSTEMS, INC.
Product Code
DQO
PMA / PMN Number
K913682
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: ONE USED SUSPECT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXAMINED VISUALLY AND THE COMPLAINT WAS CONFIRMED. THE ROTATOR WAS SEPARATED AT THE BOND BETWEEN THE COLLAR AND THE HOUSING. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO EXCEPTION DOCUMENTS. THE COMPLAINT DATA BASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE ROOT CAUSES OF THE FAILURE ARE ATTRIBUTED TO THE MANUFACTURING BONDING PROCESS. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THIS TYPE OF FAILURE. THIS SUBASSEMBLY WAS BUILT PRIOR TO IMPLEMENTATION OF THE NOTED CORRECTIVE ACTIONS. COMPLAINT DATA WILL BE MONITORED TO VERIFY EFFECTIVENESS OF CORRECTIVE ACTIONS TAKEN. EVALUATION: DEVICE HISTORY RECORD WAS REVIEWED, COMPLAINT DATA BASE WAS REVIEWED.

Description of Event or Problem · 1

THE ROTATOR ON THE STOPCOCK BROKE DURING A TRANSCATHETER HEPATIC ARTERY CHEMOEMBOLIZATION. NO HARM OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MERIT CUSTOM KIT CATHETER, INTRAVASCULAR, DIAGNOSTIC DQO MERIT MEDICAL SYSTEMS, INC. F756931

Patients

Seq Age Sex Outcome Treatment
1 CONTRAST MEDIA