FDA Adverse Event Malfunction Summary report: N

NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT

MDR report key: 20137953 · Received September 4, 2024

Report

Report Number
2435119-2024-00008
Event Type
Malfunction
Date Received
September 4, 2024
Date of Event
August 5, 2024
Report Date
February 11, 2025
Manufacturer
MISONIX, INC.
Product Code
LFL
PMA / PMN Number
K231117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ON AUGUST 05, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT (PART NUMBER 110-31-1125, LOT NUMBER 243131) DURING A SPINAL DECOMPRESSION & FUSION THAT OCCURRED ON THE SAME DAY. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. DELAY IN TREATMENT WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. SPECIFICALLY, IT WAS REPORTED DURING THE SURGEON'S CUT THAT PART OF THE BLADE TIP SNIPPED OFF, POTENTIALLY DUE TO TOO MUCH MECHANICAL FORCE. THE DEVICE HISTORY RECORD WAS REVIEWED FOR NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT (PART NUMBER 110-31-1125, LOT NUMBER 243131) IN USE AT THE TIME OF THE EVENT. THERE WERE NO DEVIATIONS FOUND DURING IN-PROCESS OR FINAL INSPECTION OF THE DEVICE. INSPECTION AND TEST RESULTS MET MISONIX SPECIFICATIONS PRIOR TO RELEASE TO COMMERCE. A REVIEW OF POST MARKET SURVEILLANCE DATA FOR THE NEXUS® SYSTEM DID NOT SHOW ANY SIGNIFICANT ADVERSE TRENDS FOR BREAKAGE OF ULTRASONIC PROBE TIPS. THE CURRENT FREQUENCY OF OCCURRENCE IS WITHIN THE FREQUENCY IN THE ORIGINAL RISK MANAGEMENT REPORT. THERE IS NO CHANGE TO THE RESIDUAL RISK OR RISK-BENEFIT RATIO OF THE DEVICE. THE RISK OF HARM IS MITIGATED BY THE NATURE IN WHICH TITANIUM ULTRASONIC BLADES BREAK DURING USE. TITANIUM DOES NOT SHATTER, SPLINTER, OR CREATE MULTIPLE FRAGMENTS THAT WOULD BE DIFFICULT TO LOCATE OR REMOVE FROM THE SURGICAL FIELD. IN MOST CASES, THE BROKEN PIECES CAN BE EASILY IDENTIFIED BY DIRECT VISUAL EXAMINATION AND REMOVED FROM THE SURGICAL FIELD. IN CASES WHERE BROKEN PIECES CANNOT BE IDENTIFIED VISUALLY, ENHANCED VISUALIZATION THROUGH THE ROUTINE USE OF LOUPE-FITTED EYEGLASSES OR MICROSCOPES CAN AID IN IDENTIFYING AND REMOVING BROKEN PIECES FROM THE SURGICAL FIELD. SURGICAL SUITES TYPICALLY HAVE ACCESS TO DIAGNOSTIC IMAGING EQUIPMENT, SUCH AS X-RAY OR FLUOROSCOPY, AND WOULD BE ABLE TO QUICKLY IDENTIFY AND LOCATE ANY FRAGMENTS NOT FOUND BY DIRECT OR ENHANCED VISUALIZATION. THE BROKEN PIECES THAT REMAIN IN THE OPERATIVE FIELD CAN THEREFORE BE FOUND AND EXTRACTED QUICKLY AND EASILY WITHOUT SIGNIFICANT DELAY IN THE SURGICAL PROCEDURE. THE INSTRUCTIONS FOR USE MANUAL (100-10-1000, REVISION K) FOR THE NEXUS® ULTRASONIC SURGICAL ASPIRATION SYSTEM CONTAINS THE FOLLOWING WARNINGS AND CAUTIONS TO PREVENT BLADE BREAKAGE. WARNING THE NEXUS® ULTRASONIC SURGICAL ASPIRATOR SYSTEM IS INTENDED TO BE USED IN VARIOUS TYPES OF INVASIVE, SURGICAL PROCEDURES. THERE MAY BE INDIRECT DANGER TO THE PATIENT SHOULD THE DEVICE FAIL DURING THE PROCEDURE. IT IS RECOMMENDED THAT THE FACILITY FOLLOWS ITS BACK-UP EQUIPMENT PROTOCOLS. WARNING ULTRASONIC PROBE TIPS CAN BREAK UNDER EXCESSIVE USE IN EXTREME CONDITIONS, E.G. WHEN CUTTING FOR EXTENDED DURATION IN TIGHT CAVITIES WITH LIMITED LATERAL MOTION. THE PROBE TIP COULD BREAK INTO TWO OR MORE FRAGMENTS WITH THE MAIN FRAGMENT REMAINING ATTACHED TO THE HANDPIECE. ALL FRAGMENTS MUST BE RETRIEVED IMMEDIATELY FROM THE SURGICAL SITE. THE FRAGMENTS SHOULD BE CHECKED TO ENSURE THAT NO FURTHER PIECES ARE MISSING. IT IS POSSIBLE THAT A FRAGMENT IS PROPELLED OUTSIDE OF THE SURGICAL CAVITY. DIAGNOSTIC IMAGING, SUCH AS X-RAY, MUST BE USED IF A FRAGMENT CANNOT BE FOUND TO CONFIRM THAT THE BROKEN PIECE IS OUTSIDE OF THE SURGICAL CAVITY. WARNING BREAKAGE OF ULTRASONIC PROBE TIPS WILL RESULT IN SHARP EDGES THAT CAN BE HARMFUL TO SOFT TISSUE EVEN WITHOUT ACTIVATION OF ULTRASOUND. PROBE TIPS CAN BEND OR DEFORM BEFORE THEY ACTUALLY BREAK. PROBE TIPS SHOWING SIGNS OF DEFORMATION OR CRACKING SHOULD BE REPLACED IMMEDIATELY SINCE PROBE TIP BREAKAGE IS OTHERWISE IMMINENT. DO NOT BEND OR TWIST THE ULTRASONIC PROBE TIPS SINCE IT REDUCES THE STRUCTURAL INTEGRITY AND CAN RESULT IN PROBE TIP BREAKAGE DURING USE. DISPOSE OF DEFORMED OR BROKEN PROBE TIPS IMMEDIATELY IN A SHARPS CONTAINER IN ACCORDANCE WITH YOUR FACILITY BIOLOGICAL HAZARDOUS WASTE PROCEDURE. WARNING DURING SYSTEM CHECK, MAKE SURE THE PROBE TIP OF THE HANDPIECE IS FREE FROM CONTACT WITH ANY OBJECT. ALLOWING CONTACT WITH THE PROBE TIP MAY RESULT IN DAMAGE AND/OR PERSONAL INJURY. CAUTION THE SYSTEM CHECK SHOULD ALWAYS BE DONE IN ADVANCE OF PREPARING PATIENT FOR SURGERY TO MINIMIZE RISK TO PATIENT IN CASE OF SYSTEM MALFUNCTION. REFER TO SECTION 7.11 - SYSTEM CHECK. CAUTION LOOSE PROBE TIP/TISSUE CONTACT UPON AN INITIAL BONE INCISION CAN CAUSE A THIN PROBE TIP TO RESONATE NOT ONLY LONGITUDINALLY BUT ALSO TRANSVERSELY. THIS CAN CAUSE A THIN PROBE TIP TO BREAK. IT IS NECESSARY TO ENGAGE BONE ACTIVELY AND WITH A MINIMAL PROBE TIP PRESSURE GREATER THAN ZERO IN ORDER TO PREVENT THE SHATTERING. CAUTION CONTACT OF THE ULTRASONIC PROBE TIP OR THE EXPOSED EXTENSION WITH METAL, SURGICAL INSTRUMENTS OR OTHER OBJECTS DURING ULTRASOUND USE MUST BE AVOIDED. SUCH CONTACT CAN DAMAGE THE ULTRASONIC COMPONENTS VERY EASILY AND MAY RESULT IN COMPROMISED PERFORMANCE, INCLUDING FAILURE. DISCARD ANY EXTENSIONS OR PROBE TIPS THAT SHOW SIGNS OF DAMAGES LIKE GOUGES, NICKS OR FRACTURES. EXTERNAL ASPIRATION MAY BE USED BUT IT IS RECOMMENDED THAT A PLASTIC SUCTION PROBE TIP SHOULD BE USED WHEN IN PROXIMITY WITH THE PROBE TIP. CAUTION THE NEXUS® SYSTEM SHOULD BE FULLY TESTED AND INSPECTED PRIOR TO EACH PROCEDURE. THE CONSOLE, FOOTSWITCH, HANDPIECES, ALL CABLES AND ACCESSORIES SHOULD BE EXAMINED FOR PROPER APPEARANCE AND CONDITION. THE CONSOLE MONITORS THE ULTRASONIC OUTPUT AT ALL TIMES AND ALERTS IN CASES OF OVERLOAD OR MALFUNCTION OF THE VIBRATING ELEMENTS (HANDPIECE, EXTENSION AND ULTRASONIC TIP). A NOTIFICATION IS DISPLAYED TOGETHER WITH AN AUDIBLE INDICATOR AS LONG AS THE FOOTSWITCH IS DEPRESSED. TIP OVERLOAD CAN OCCUR DURING HARD TISSUE REMOVAL WHEN APPLYING EXCESSIVE TIP PRESSURE OR FACING STRONG TISSUE RESISTANCE, E.G. FROM THICK CORTICAL BONE. THIS CAN LEAD TO STALLING OF THE ULTRASONIC TIP. FOLLOW THE CORRECTIVE ACTION BELOW. TABLE 8.2 CONTAINS STEPS FOR THE USER TO TAKE SHOULD A MECHANICAL LIMIT ALERT APPEAR DURING USE AS WELL AS RECOMMENDED CORRECTIVE ACTIONS TO TAKE. THE POSSIBLE CAUSES LISTED FOR A MECHANICAL LIMIT ALERT ARE: TIP OVERLOAD, LOOSE OR DAMAGED COMPONENT, DEFECTIVE HANDPIECE. NUMBERS 1 AND 2 HAVE CORRECTIVE ACTIONS LISTED THAT CAN BE PERFORMED BY THE USER, HOWEVER IF THE LISTED CORRECTIVE ACTIONS ARE FOLLOWED AND THE ALERT CONTINUES, THE HANDPIECE MAY NEED TO BE REPLACED. IN THE EVENT THAT THE HANDPIECE NEEDS TO BE REPLACED, THE FACILITY SHOULD FOLLOW THEIR BACK-UP EQUIPMENT PROTOCOLS AS LISTED IN THE GENERAL SAFETY REQUIREMENTS. THE DISPOSABLE HAS BEEN RETURNED BUT HAS NOT YET BEEN EVALUATED TO DETERMINE POTENTIAL ROOT CAUSE. A FOLLOW UP REPORT WILL BE ISSUED ONCE EVALUATION OF THE DISPOSABLE IS COMPLETE.

Additional Manufacturer Narrative · 0

THE NEXUS® BONESCALPEL® 25MM, BLUNT BLADE + IRRIGATION TUBING KITS (PART NUMBER 110-31-1125) IS A STERILE KIT CONTAINING MULTIPLE COMPONENT PARTS. THE 25MM BLADE (PART NUMBER E4003ABC25) RETURNED SHOWS A BLADE FRACTURED APPROXIMATELY .15" FROM THE PROXIMAL END OF THE SLOT ON ONE LEG LATERALLY TO ITS OUTSIDE EDGE, TO THE DISTAL END OF THE SLOT THROUGH TO THE TIP, SUCH THAT THE MAJORITY OF ONE HALF OF THE BLADE HAS FULLY DETACHED. POSSIBLE CAUSES OF FAILURE INCLUDE INSUFFICIENT IRRIGATION AND IMPROPER TIP ENGAGEMENT IN HARD TISSUE. IRRIGATION DAMPENS UNWANTED FLEXURAL MOTION, WHICH IS INHERENTLY AMPLIFIED IN LONGER BLADES. IF ULTRASOUND WAS ACTIVATED FOR AN EXTENDED PERIOD WITHOUT THE BENEFIT OF THE DAMPING PROVIDED BY IRRIGATION, THE FATIGUE LIMIT OF THE BLADE IS QUICKLY REACHED, RESULTING IN FAILURE. SIMILARLY, IF THE TIP WAS ENGAGED IN HARD TISSUE AND SUBJECTED TO TWISTING OR BENDING FORCES, THE COMBINATION OF CYCLIC ULTRASONIC STRESSES AND STRESSES FROM APPLIED FORCES COULD EXCEED THE MATERIAL FATIGUE LIMIT. THE COMMENT REGISTERED WITH THE COMPLAINT, "SEEMED TO BE FROM TOO MUCH MECHANICAL FORCE", WOULD SEEM TO CORROBORATE SUCH A CONCLUSION. THE INVESTIGATION HAS BEEN CONCLUDED.

Description of Event or Problem · 0

ON AUGUST 05, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT (PART NUMBER 110-31-1125, LOT NUMBER 243131) DURING A SPINAL DECOMPRESSION & FUSION THAT OCCURRED ON THE SAME DAY. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. DELAY IN TREATMENT WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. SPECIFICALLY, IT WAS REPORTED DURING THE SURGEON'S CUT THAT PART OF THE BLADE TIP SNIPPED OFF, POTENTIALLY DUE TO TOO MUCH MECHANICAL FORCE.

Description of Event or Problem · 0

ON AUGUST 05, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT (PART NUMBER 110-31-1125, LOT NUMBER 243131) DURING A SPINAL DECOMPRESSION & FUSION THAT OCCURRED ON THE SAME DAY. A SERIOUS INJURY TO THE PATIENT OR USER WAS NOT REPORTED. DELAY IN TREATMENT WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. SPECIFICALLY, IT WAS REPORTED DURING THE SURGEON'S CUT THAT PART OF THE BLADE TIP SNIPPED OFF, POTENTIALLY DUE TO TOO MUCH MECHANICAL FORCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1315927 NEXUS® BONESCALPEL® 25MM BLUNT BLADE AND IRRIGATION TUBING KIT 25MM BLUNT BLADE AND IRRIGATION TUBING KIT LFL MISONIX, INC. 110-31-1125 243131

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other