HIRES¿ ULTRA IMPLANT
Report
- Report Number
- 3006556115-2024-01288
- Event Type
- Malfunction
- Date Received
- September 4, 2024
- Date of Event
- July 29, 2024
- Report Date
- August 14, 2024
- Manufacturer
- ADVANCED BIONICS, LLC
- Product Code
- MCM
- UDI-DI
- 07630016855696
- PMA / PMN Number
- P960058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
ADDITIONAL INFORMATION: SECTION D.9 DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
ADVANCED BIONICS CONSIDERS THE INVESTIGATION INTO THIS REPORTABLE EVENT AS CLOSED. THE EXTERNAL VISUAL INSPECTION REVEALED THE ELECTRODE WAS SEVERED AND SILICONE DAMAGE WAS OBSERVED ON THE TOP AND BOTTOM COVER OF THE DEVICE. THESE ARE BELIEVED TO HAVE OCCURRED DURING REVISION SURGERY. THE DEVICE PASSED PHOTOGRAPHIC IMAGING INSPECTION. SYSTEM LOCK WAS VERIFIED. THE CONDITION OF THE ELECTRODE PREVENTED AN ELECTRICAL TEST FROM BEING PERFORMED. THE DEVICE PASSED SOME OF THE ELECTRICAL TESTS PERFORMED. THE FAILURE OF THIS DEVICE IS ATTRIBUTED TO AN ELECTRODE SHORT IN THE ELECTRODE POCKET. A CORRECTIVE ACTION WAS IMPLEMENTED. THIS VERSION OF THE HIRES ULTRA DEVICE IS NO LONGER DISTRIBUTED. THIS IS THE FINAL REPORT. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
THE RECIPIENT REPORTEDLY EXPERIENCED DECREASED PERFORMANCE. A REVIEW OF THE RECIPIENT¿S TEST DATA INDICATED IMPEDANCE ISSUES. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1105331 | HIRES¿ ULTRA IMPLANT | COCHLEAR IMPLANT | MCM | ADVANCED BIONICS, LLC | CI-1600-05 | NA | 07630016855696 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |