FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM

MDR report key: 2013663 · Received February 4, 2011

Report

Report Number
2242352-2011-00023
Event Type
Malfunction
Date Received
February 4, 2011
Date of Event
January 5, 2011
Report Date
January 6, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED WITHOUT THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE DELIVERY TUBE. THE SEAL WAS ROLLED BUT EXTENDED MORE THAN HALFWAY OUTSIDE THE TUBE. THE SEAL WAS CRACKED CLOSER TO THE CENTER RING. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THESE FINDINGS, THE REPORTED FAILURE "SEAL CRACKED DURING LOADING" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE SEAL CRACKED DURING LOADING. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.