FDA Adverse Event Injury Summary report: N

ENDOVIVE INITIAL PLACEMENT PEG KIT, DIRECT PEJ KIT, AND PEG SAFETY KIT

MDR report key: 2013612 · Received March 10, 2011

Report

Report Number
3005099803-2011-00678
Event Type
Injury
Date Received
March 10, 2011
Date of Event
February 18, 2011
Report Date
February 18, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
KNT
PMA / PMN Number
K031538
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT WAS UNABLE TO PROVIDE THE SUSPECT DEVICE LOT NUMBER; THEREFORE, THE LOT EXPIRATION AND DEVICE MANUFACTURE DATES ARE UNKNOWN. (B)(4) - INCISION MADE LARGER. (B)(4) - THE REPORTED EVENT OF DIFFICULTY PULLING THE PEG TUBE THROUGH THE INCISION SITE. THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION, AS IT IS IMPLANTED; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PULL METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6) 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE NURSE REPORTED THEY HAD EXTREME DIFFICULTY PULLING THE PEG THROUGH THE INCISION SITE. THE BOSTON SCIENTIFIC SALES REPRESENTATIVE CONFIRMED THAT THE INITIAL INCISION SITE WAS OF PROPER LENGTH. THE RESISTANCE WAS FELT AT THE POINT WHERE THE CONE SHAPE ON THE END OF THE PEG MEETS THE SOFT PLASTIC. THE PHYSICIAN EXTENDED THE INCISION SITE AND WAS ABLE TO PLACE THIS DEVICE. THE PATIENT'S CONDITION WAS REPORTED TO BE FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOVIVE INITIAL PLACEMENT PEG KIT, DIRECT PEJ KIT, AND PEG SAFETY KIT TUBES, GASTROINTESTINAL (AND ACCESSORIES) KNT BOSTON SCIENTIFIC - SPENCER M00566461

Patients

Seq Age Sex Outcome Treatment
1 56 YR Other