FDA Adverse Event Malfunction Summary report: N

SMARTTOUCH

MDR report key: 20135452 · Received September 4, 2024

Report

Report Number
1000165971-2024-00882
Event Type
Malfunction
Date Received
September 4, 2024
Date of Event
August 8, 2024
Report Date
October 25, 2024
Manufacturer
MICROPORT CRM S.R.L.
Product Code
KRG
UDI-DI
08031527021140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REVIEW OF PROVIDED DATA CONFIRMED THE REPORTED ISSUE ON (B)(6) 2024. THE PROVIDED EXPERT FILES SHOW THAT THE FOLLOW UP OF THE IMPLANTABLE PACEMAKER S/N (B)(6) STARTED ON (B)(6) AT 12:07 DID NOT END NORMALLY, AND THAT THE BATTERY WAS REMOVED AROUND 12:30. THE PROGRAMMER BECAME STUCK BEFORE THE END OF THE SESSION. THE ISSUE IS LINKED TO INTERNAL MANAGEMENT OF THE PROGRAMMER SOFTWARE (REPLY). THIS ISSUE WILL BE TRACED FOR FUTURE IMPROVEMENTS. NO GENERAL MALFUNCTION IS SUSPECTED ON THE SUBJECT DEVICE. THIS CASE IS RETAINED AND UTILIZED FOR TREND PURPOSES.

Description of Event or Problem · 0

REPORTEDLY, I RECEIVED A CALL TODAY FROM (B)(6). THEY USE A SMARTTOUCH PROGRAMMING SYSTEM WITH SOFTWARE VERSION 3.16. THERE WAS A SCREEN FREEZE ISSUE. THE PROBLEM WAS RESOLVED BY REMOVING THE BATTERY.

Description of Event or Problem · 0

REPORTEDLY, I RECEIVED A CALL TODAY FROM LKH GRAZ WEST. THEY USE A SMARTTOUCH PROGRAMMING SYSTEM WITH SOFTWARE VERSION 3.16. THERE WAS A SCREEN FREEZE ISSUE. THE PROBLEM WAS RESOLVED BY REMOVING THE BATTERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
848314 SMARTTOUCH PROGRAMMER, PACEMAKER KRG MICROPORT CRM S.R.L. SMARTTOUCH SOFTWARES MODULES 08031527021140

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown