ACTIV POWDER GP CEMENT
Report
- Report Number
- 1032227-2011-00001
- Event Type
- Malfunction
- Date Received
- February 3, 2011
- Date of Event
- January 7, 2011
- Report Date
- February 3, 2011
- Manufacturer
- BRASSELER USA
- Product Code
- EMA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- DENTIST
Narratives
PRODUCT WAS RETURNED FROM CUSTOMER AND WAS EVALUATED BY COMPANY PERSONNEL. THE PRODUCT REFERENCED IN THE COMPLAINT IS PART OF A TWO-PART CEMENT. THE POWDER IS MIXED WITH A LIQUID TO PRODUCE THE CEMENT USED IN ENDODONTIC PROCEDURES TO SEAL TOOTH CANAL TUBULARS AND ACT AS AN ADHERENT FOR THE ROOT CANAL FILLER (I..E. GUTTA PERCHA). THE PRODUCT WAS MIXED WITH THE LIQUID (ALSO RETURNED FROM THE CUSTOMER) AND ALLOWED TO SET UP. WHEN THE PRODUCT WAS MIXED ACCORDING TO DIRECTIONS AS A 'CLUMP' OR 'BALL' (SMALL SURFACE AREA), THE SET UP TIME WAS BETWEEN 11.5 MINUTES AND 45 MINUTES. WHEN THE PRODUCT WAS MIXED ACCORDING TO DIRECTIONS AS A SMEAR (GREATER SURFACE AREA), THE PRODUCT SET UP TIME WAS BETWEEN 11.5 AND 15 MINUTES. INSTRUCTIONS FOR USE INDICATE THE PRODUCT HAS A WORKING TIME OF 11.5 MINUTES FROM COMPLETION OF THE MIXING OF COMPONENTS UNTIL THE SEALER IS SET. THIS TIME CAN BE EXTENDED TO 20 MINUTES BY MIXING ON A CHILLED GLASS SLAB. THE REMAINING PRODUCT WAS FORWARDED TO THE THIRD PARTY MANUFACTURER FOR ADDITIONAL INVESTIGATION. CONCLUSION: NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE DENTAL OFFICE ADVISED THAT WHEN THE DENTIST USED THE POWDER AND LIQUID, THE SEALER DID NOT SET UP WHEN MIXED. THERE WAS NO PATIENT INJURY REPORTED AS A RESULT OF THIS MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACTIV POWDER GP CEMENT | EMA | BRASSELER USA | ACTIV GP | 1705145-15B0809 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |