FDA Adverse Event Injury Summary report: N

HILL ROM

MDR report key: 2013482 · Received March 7, 2011

Report

Report Number
MW5019696
Event Type
Injury
Date Received
March 7, 2011
Date of Event
January 25, 2011
Report Date
March 7, 2011
Manufacturer
HILL ROM
Product Code
FNL
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
ND, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PT'S MOTHER SAW SPARKS COMING FROM BED. NURSING STAFF RESPONDED. PT WAS IN GRANDFATHER'S LAP IN CHAIR. NO FIRE WAS SEEN, BUT ROOM SMELLED "HOT." STAFF ASKED GRANDFATHER TO PICK PT UP AND CARRY OUT OF ROOM. PT ROOM DOOR WAS SHUT. FIRE ALARM PULLED AND CODE RED CALLED TO SWITCHBOARD. FACILITIES MANAGEMENT STAFF ARRIVED. ROOM DOOR OPENED AND SMOKE NOTED. UNPLUGGED THE BED. BED INSPECTED. FLAMES NOTED, SO FIRE EXTINGUISHER USED TO PUT OUT FLAMES. SAFETY OFFICER AND FIRE DEPARTMENT ARRIVED. RECEIVED UPDATE. OTHER PTS ALLOWED TO REMAIN IN ROOMS. FIREMEN REMOVED BED FROM ROOM AND AREA. PER BED EXAM, FIRE ORIGINATED NEAR ACTUATOR MOTOR CAPACITATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HILL ROM PATIENT BED FNL HILL ROM 850 CENTRA

Patients

Seq Age Sex Outcome Treatment
1 3 YR Disability