FDA Adverse Event Malfunction Summary report: N

BILIARY STENT, BILIARY STENT SYSTEM, BILIARY ENDOPROSTHESIS

MDR report key: 2013430 · Received March 10, 2011

Report

Report Number
3005099803-2011-00636
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FGE
PMA / PMN Number
K965147
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE REVEALED THAT THE STENT WAS STILL ATTACHED VIA THE UNBROKEN SUTURE. THE PUSH CATHETER WAS KINKED NEAR THE HUB. THE SUTURE HOLE WAS TORN ON THE DISTAL END OF THE PUSH CATHETER. THE GUIDE CATHETER WAS STRETCHED AND BROKEN. DURING THE INVESTIGATION, THE SUTURE WAS CUT AND THE STENT RELEASED. THE GUIDE CATHETER REVEALED SUTURE LIGATURE MARKS. THERE WAS NO DAMAGE TO THE STENT. BASED ON THE CONDITION OF THE DEVICE AND THE DETAILS OF THE EVENT, THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS OPERATIONAL CONTEXT. THE DEVICE HISTORY RECORD REVIEW CONFIRMS THAT THE DEVICE MET ALL MANUFACTURING SPECIFICATIONS. A LOT HISTORY SEARCH WAS PERFORMED AND FOUND NO OTHER COMPLAINTS AGAINST THE SPECIFIED LOT.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED; HOWEVER THE EVALUATION HAS NOT YET BEEN COMPLETED. IF THERE IS ANY FURTHER RELEVANT INFORMATION, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT SYSTEM WAS USED DURING A STENTING PROCEDURE IN THE COMMON BILE DUCT OF A FEMALE PATIENT (PATIENT AGE AND WEIGHT ARE UNKNOWN). DURING THE PROCEDURE, RESISTANCE WAS MET AS THE GUIDE CATHETER WAS RETRACTED FOR STENT DEPLOYMENT. THE GUIDE CATHETER STRETCHED, BROKE, AND THE STENT WAS UNABLE TO BE DEPLOYED. THE BROKEN GUIDE CATHETER REMAINED WITHIN THE PUSH CATHETER ALLOWING THE DEVICE TO BE REMOVED IN ONE PIECE. THE PROCEDURE WAS ENDED WITHOUT A STENT AS THE PHYSICIAN WAS UNABLE TO CROSS THE STRICTURE A SECOND TIME WITH THE OLYMPUS VISIGLIDE GUIDEWIRE. THERE WERE NO REPORTED PATIENT COMPLICATIONS. THE PATIENT WAS REPORTED TO BE IN GOOD CONDITION AFTER THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT SYSTEM WAS USED DURING A STENTING PROCEDURE IN THE COMMON BILE DUCT OF A FEMALE PATIENT (PATIENT AGE AND WEIGHT ARE UNKNOWN). DURING THE PROCEDURE, RESISTANCE WAS MET AS THE GUIDE CATHETER WAS RETRACTED FOR STENT DEPLOYMENT. THE GUIDE CATHETER STRETCHED, BROKE, AND THE STENT WAS UNABLE TO BE DEPLOYED. THE BROKEN GUIDE CATHETER REMAINED WITHIN THE PUSH CATHETER ALLOWING THE DEVICE TO BE REMOVED IN ONE PIECE. THE PROCEDURE WAS ENDED WITHOUT A STENT AS THE PHYSICIAN WAS UNABLE TO CROSS THE STRICTURE A SECOND TIME WITH THE OLYMPUS VISIGLIDE GUIDEWIRE. THERE WERE NO REPORTED PATIENT COMPLICATIONS. THE PATIENT WAS REPORTED TO BE IN GOOD CONDITION AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BILIARY STENT, BILIARY STENT SYSTEM, BILIARY ENDOPROSTHESIS CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - SPENCER M00539210 13592683

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: OLYMPUS VISIGLIDE