FDA Adverse Event
Malfunction
Summary report: N
CESAR-OMCP-VISUB (HM3000)
MDR report key: 2013415
·
Received February 2, 2011
Report
- Report Number
- 3003768277-2011-00090
- Event Type
- Malfunction
- Date Received
- February 2, 2011
- Report Date
- January 5, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- IZF
- PMA / PMN Number
- K925302
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE INVESTIGATION SHOWED THAT THE FIELD SERVICE ENGINEER (FSE) FOUND THE GEO CEASER EXTENSION BOARD WAS BURNT. THE BOARD WAS REPLACED AND OPERATION VERIFIED. THE SYSTEM'S END OF SERVICE WAS 2010. THE REPLACEMENT BOARD WAS FROM THE HOSPITAL STOCK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SCREEN WILL NOT BOOT NO FLUORO AND THERE IS A BURNING SMELL IN THE EQUIPMENT ROOM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CESAR-OMCP-VISUB (HM3000) | IZF | PHILIPS HEALTHCARE | 72239 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |