FDA Adverse Event Malfunction Summary report: N

CESAR-OMCP-VISUB (HM3000)

MDR report key: 2013415 · Received February 2, 2011

Report

Report Number
3003768277-2011-00090
Event Type
Malfunction
Date Received
February 2, 2011
Report Date
January 5, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
IZF
PMA / PMN Number
K925302
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION SHOWED THAT THE FIELD SERVICE ENGINEER (FSE) FOUND THE GEO CEASER EXTENSION BOARD WAS BURNT. THE BOARD WAS REPLACED AND OPERATION VERIFIED. THE SYSTEM'S END OF SERVICE WAS 2010. THE REPLACEMENT BOARD WAS FROM THE HOSPITAL STOCK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SCREEN WILL NOT BOOT NO FLUORO AND THERE IS A BURNING SMELL IN THE EQUIPMENT ROOM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CESAR-OMCP-VISUB (HM3000) IZF PHILIPS HEALTHCARE 72239

Patients

Seq Age Sex Outcome Treatment
1