FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2013399 · Received March 10, 2011

Report

Report Number
3005075853-2011-00972
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
November 30, 2010
Report Date
December 14, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). JAWS. THE ANALYSIS RESULTS OF DEVICE (A) FOUND THAT IT WAS RETURNED EMPTY AND WITH THE LOCK OUT MECHANISM BROKEN AS IT WAS FIRED THROUGH. THE INSTRUMENT IS DESIGNED TO LOCK OUT AFTER ALL THE CLIPS HAVE BEEN FIRED; THEREFORE A POTENTIAL CAUSE OF THE CUSTOMER REPORTED EXPERIENCE IS THE FIRING OF ALL OF THE CLIPS AND THE INSTRUMENT "WOULD NOT FIRE" (ACTIVATION OF THE LOCK OUT MECHANISM). THE INSTRUMENT HAS AN ORANGE INDICATOR THAT APPEARS ON THE TOP OF THE HANDLE AS A REFERENCE FOR THE USER AS TO THE QUANTITY OF CLIPS REMAINING. THE ANALYSIS RESULTS FOUND THAT THE DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. BATCH # G9L57X MFG DATE: 9/30/2010 EXP DATE: 08/30/2015. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, BOTH DEVICES JAMMED. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. NO ADVERSE CONSEQUENCES REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4U206

Patients

Seq Age Sex Outcome Treatment
1