30K FSI-SLI-FG-10S INS,PKD
Report
- Report Number
- 2424472-2024-00392
- Event Type
- Malfunction
- Date Received
- September 4, 2024
- Date of Event
- August 15, 2024
- Report Date
- September 4, 2024
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- UDI-DI
- D003820051
- PMA / PMN Number
- K052334/K970
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. INVESTIGATION RESULTS: RETURN QTY. 1, 22321, 30K FSI-SLI-10S-GF, 00087051 BUL 30K. TEST METHOD / GP005-WI02 INDUCTANCE: GAUGE ID (B)(4) DUE: (B)(6) 2024 RESULT 3.05. MEETS SPEC DOES NOT MEET SPEC. N/A TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC. N/A STACK CONDITION MEETS SPEC DOES NOT MEET SPEC. N/A TESTED ON: G130 GAGE ID (B)(4) DUE DATE: ON (B)(6) 2024 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC. N/A SPRAY PATTERN: - MEETS SPEC DOES NOT MEET SPEC. N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC, N/A GRIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A OTHER VISUAL OBSERVATION: INSERT TIP BROKEN AT THE EDM HOLE. ALLEGED EVENT: CONFIRMED - NOT CONFIRMED.
IN THIS EVENT IT IS REPORTED THAT A 30K FSI-SLI-FG-10S INS,PKD TIP BROKE AND PATIENT SWALLOWED THE BROKEN TIP. PATIENT WAS SENT TO EMERGENCY ROOM TO LOCATE TIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 870406 | 30K FSI-SLI-FG-10S INS,PKD | SCALER, ULTRASONIC | ELC | DENTSPLY LLC | 00087051 | D003820051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |