FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 20133513 · Received September 4, 2024

Report

Report Number
3003442380-2024-23792
Event Type
Malfunction
Date Received
September 4, 2024
Date of Event
July 31, 2024
Report Date
August 7, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016651
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1961118- MDR 3003442380-2024-23792- DEVICE 1 OF 2.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(6). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED CANNULA ISSUES WITH TWO INFUSION SETS. THE CANNULAS WERE KINKED. THE EVENT OCCURRED ON 31-JUL-2024. PATIENT NOTICED SYMPTOMS WITHIN 3 HOURS OF INSERTION. THE INSERTION OF SITE WAS THE ABDOMEN. PATIENT REGULARLY ROTATED SITE LOCATION. PATIENT CONFIRMED THAT THE INTRODUCER NEEDLE WAS AHEAD OF THE CANNULA PRIOR TO INSERTION. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
870382 AUTOSOFT XC UNO INSET I 60/9 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001681 UNKNOWN 05705244016651

Patients

Seq Age Sex Outcome Treatment
1 82 YR Female