AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-23874
- Event Type
- Malfunction
- Date Received
- September 4, 2024
- Date of Event
- August 4, 2024
- Report Date
- November 15, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED COMMON DEVICE NAME UNDER D2, MODEL NUMBER, SERIAL NUMBER, PRIMARY UNIQUE DEVICE IDENTIFIER (UDI) NUMBER. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF RECORD (B)(4) DOES NOT REQUIRE A TYPE 2 REPORTABLE COMPLAINT TO INITIATE A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS CREATED AND SUBSEQUENTLY CLOSED TO DOCUMENT THE DHR REVIEW, WHICH WAS NECESSARY TO ADVANCE THE PARENT RECORD TO THE REVIEW AND SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6002595, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6002595 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 106 AND MANUFACTURED IN THE LINE 5 ON 03/AUG/2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 3G02740 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 55 AND MANUFACTURED IN THE MACHINE SC04, ON 31/JUL/2023, WITH A TOTAL OF (B)(4) UNITS THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 3G04238 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI)WORK INSTRUCTION (WI) VERSION 55 AND MANUFACTURED IN THE MACHINE SC04, ON 01/AUG/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: ONE NON CONFORMANCE (NC) RAISED DURING STERILIZATION PROCESS UNRELATED TO COMPLAINT CODE, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT ENCOUNTERED INFUSION SET OCCLUSION IN TUBING. THE BLOOD GLUCOSE WAS REPORTED HIGH AND TREATED WITH BOLUS VIA PUMP. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442428 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001680 | 6002595 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female |