CONTINU-FLO SOLN. SET, 3 LUERACTIVATED VALVES

MDR report key: 2013332 · Received March 10, 2011

Report

Report Number
6000001-2011-01761
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
February 1, 2011
Report Date
February 22, 2011
Manufacturer
BAXTER HEALTHCARE
Product Code
FPA
PMA / PMN Number
K961225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION ON 03/08/2011. ONE ACTUAL SAMPLE WAS RECEIVED FOR THE REPORTED CONDITION OF A CUT NEAR A CHECK VALVE. A VISUAL INSPECTION DETERMINED ALL COMPONENTS WERE PRESENT, IN THE RIGHT POSITION AND COMPLETE. THE SAMPLE WAS THEN SUBMITTED FOR AN UNDERWATER PRESSURE TEST AND THE REPORTED CONDITION WAS CONFIRMED. ALTHOUGH THE REPORTED CONDITION WAS CONFIRMED, THE ROOT CAUSE OF THIS CONDITION COULD NOT IDENTIFIED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BAXTER A CONTINU-FLO SOLUTION SET IN WHICH A "CUT NEAR A CHECK VALVE" WAS DETECTED. THE REPORTER STATED THAT IT APPEARED THE CUT COULD HAVE POSSIBLY OCCURRED DURING PACKAGING. THE CONDITION WAS IDENTIFIED DURING SET-UP BEFORE PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.