FDA Adverse Event Malfunction Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 2013314 · Received March 10, 2011

Report

Report Number
2955842-2011-00067
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
February 9, 2011
Report Date
February 9, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER ACCESSORY HAS NOT BEEN RETURNED FOR EVALUATION, HOWEVER, THE MONOPOLAR CURVED SCISSORS INSTRUMENT USED WITH THE TIP COVER DURING THE PROCEDURE WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING FOUND NO CHAR MARKS OR SIGNS OF ARCING ON THE INSTRUMENT. ELECTRICAL CONTINUITY PASSED AND NO DAMAGE WAS FOUND. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, WHILE USING THE MONOPOLAR CURVED SCISSORS INSTRUMENT WITH THE TIP COVER ACCESSORY INSTALLED, ARCING WAS OBSERVED. NO PATIENT INJURY, HARM OR ADVERSE OUTCOME WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180-10 C10344

Patients

Seq Age Sex Outcome Treatment
1 55 YR DA VINCI S SYSTEM, INSTRUMENTS, ACCESSORIES, & ES