FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER

MDR report key: 2013252 · Received March 10, 2011

Report

Report Number
3005075853-2011-00968
Event Type
Malfunction
Date Received
March 10, 2011
Report Date
December 3, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). JAWS. THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. FURTHER INSPECTION FOUND THAT THE JAWS HAD BECOME YIELDED CAUSING THE CLIPS TO BE EJECTED AND MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. NO DETAILS OF THE EVENT ARE AVAILABLE. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK G4TX39

Patients

Seq Age Sex Outcome Treatment
1