LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2011-00968
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Report Date
- December 3, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). JAWS. THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. FURTHER INSPECTION FOUND THAT THE JAWS HAD BECOME YIELDED CAUSING THE CLIPS TO BE EJECTED AND MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT.
IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. NO DETAILS OF THE EVENT ARE AVAILABLE. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | UNK | G4TX39 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |