LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2011-00966
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- February 14, 2011
- Report Date
- February 15, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOCKOUT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME ADDITIONAL INFORMATION.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY PROCEDURE, THE JAW DID NOT OPEN AFTER FIRING. THE DEPLOYED CLIPS WERE FORMED PROPERLY. THE DEVICE WAS RETRIEVED BY CLIPPING THE BLOOD VESSEL WITH THE OTHER NEW DEVICE AND THEN CUT BETWEEN THE DEPLOYED CLIPS. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE DOCTOR COMMENTED THAT IT HAD A DIFFICULTY IN FIRING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |