ASR ACETABULAR IMPLANT 50
Report
- Report Number
- 1818910-2011-02064
- Event Type
- Injury
- Date Received
- February 24, 2011
- Date of Event
- June 24, 2010
- Report Date
- January 4, 2013
- Manufacturer
- DEPUY INTERNATIONAL, LTD.
- Product Code
- KWA
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. ALTHOUGH THE ROOT CAUSE CANNOT BE DETERMINED, IT IS KNOWN THAT THE ASR PLATFORM WAS VOLUNTARILY RECALLED IN AUGUST OF 2010 FOLLOWING AN HHE (HEALTH HAZARD EVALUATION). FURTHER ANALYSIS REGARDING THE ASR PRODUCT FAMILY WILL BE DOCUMENTED, AS DETERMINED PERTINENT, IN (B)(4). BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR FURTHER CORRECTIVE ACTION IS NOT INDICATED AT THIS TIME.
DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
ADDITIONAL NARRATIVE: PATIENT HARMS/ IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
SURGEON CONFIRMATION FORM RECEIVED. SCF INDICATES THAT THE PATIENT WAS REVISED TO ADDRESS PAIN, NOISE, DISLOCATION AND ALVAL/SOFT TISSUE REACTION.
MEDICAL RECORD RECEIVED. THE RECORD INDICATES THAT THE PATIENT WAS EXPERIENCING SEVERE SOFT TISSUE AND BONY DAMAGE DUE TO THE REACTION TO THE METAL-ON-METAL IONS PRIOR TO BEING REVISED. THE PATIENT'S ABDUCTORS WERE COMPLETELY DESTROYED AND PULLED OFF LEAVING A BARE TROCHANTER. THERE WAS NO BACTERIAL GROWTH FROM THE CULTURES TAKEN. THE HISTOPATHOLOGY REPORT WAS CONSISTENT WITH ALVAL TYPE OF REACTION TO METAL-ON-METAL IONS.
THE PT HAS HAD AN ASR REVISION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASR ACETABULAR IMPLANT 50 | 87KWA | KWA | DEPUY INTERNATIONAL, LTD. | NA | 2263824 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Required Intervention |