FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUP

MDR report key: 2013154 · Received February 24, 2011

Report

Report Number
1818910-2011-02066
Event Type
Injury
Date Received
February 24, 2011
Date of Event
June 24, 2010
Report Date
January 4, 2013
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. ALTHOUGH THE ROOT CAUSE CANNOT BE DETERMINED, IT IS KNOWN THAT THE ASR PLATFORM WAS VOLUNTARILY RECALLED IN AUGUST OF 2010 FOLLOWING AN HHE (HEALTH HAZARD EVALUATION). FURTHER ANALYSIS REGARDING THE ASR PRODUCT FAMILY WILL BE DOCUMENTED, AS DETERMINED PERTINENT, IN (B)(4). BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR FURTHER CORRECTIVE ACTION IS NOT INDICATED AT THIS TIME.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

(B)(4). DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN. THE INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS WAS CONDUCTED UNDER MDD CAPA-(B)(4). ONGOING POST MARKET SURVEILLANCE IS CONDUCTED PER OUR PROCEDURES FOR THIS PRODUCT. DEVICE HISTORY LOT: NULL. DEVICE HISTORY BATCH: NULL. DEVICE HISTORY REVIEW: NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). ADDITIONAL NARRATIVE: PATIENT HARMS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Description of Event or Problem · 1

REASON(S) FOR REVISION: UNKNOWN.

Description of Event or Problem · 1

CLAIMSUITE RECEIVED. CLAIMSUITE ALLEGE THAT PATIENT WAS REVISED FOR PAIN.

Description of Event or Problem · 1

SURGEON CONFIRMATION FORM RECEIVED. SCF INDICATES THAT THE PATIENT WAS REVISED TO ADDRESS PAIN, NOISE AND ALVAL/SOFT TISSUE REACTION.

Description of Event or Problem · 1

MEDICAL RECORD RECEIVED. THE RECORD INDICATES THAT PRIOR TO REVISION, THE PATIENT HAD AN UNDERLYING SOFT TISSUE MASS (PSEUDOTUMOR). PATIENT COMPLAINED OF GROIN PAIN AND PAIN REFERRED TO THE KNEE. THE PATIENT WAS USING CRUTCHES. IT WAS ALSO INDICATED THAT DURING THE REVISION ON THE CONTRALATERAL HIP, THE HISTOPATHOLOGY REPORT ON THAT HIP WAS CONSISTENT WITH ALVAL TYPE OF REACTION TO METAL-ON-METAL IONS.

Description of Event or Problem · 1

THE PT HAS HAD AN ASR REVISION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUP 87KWA KWA DEPUY INTERNATIONAL, LTD. NA 2370103

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention