FDA Adverse Event Injury Summary report: N

ARTICULEZE BALL 28 +1.5 GR

MDR report key: 2013084 · Received February 24, 2011

Report

Report Number
1818910-2011-02699
Event Type
Injury
Date Received
February 24, 2011
Date of Event
January 25, 2011
Report Date
January 25, 2011
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JDI
PMA / PMN Number
K883460
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS GROIN AND HIP PAIN, OSTEOLYSIS, AND POLYWEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARTICULEZE BALL 28 +1.5 GR 87JDI JDI DEPUY ORTHOPAEDICS, INC. NA SB6A11

Patients

Seq Age Sex Outcome Treatment
1 39 YR Required Intervention