FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2012917 · Received March 4, 2011

Report

Report Number
2531779-2011-01351
Event Type
Malfunction
Date Received
March 4, 2011
Date of Event
February 3, 2011
Report Date
February 3, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE CARTRIDGE WAS RETUNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT # B201582 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FORM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE PT NOTICED LEAKAGE IN THE CARTRIDGE COMPARTMENT AND INSIDE THE PLUNGER OF THE CARTRIDGE. THE PT STATED THAT THERE WAS CONDENSATION ON THE INSIDE OF THE CARTRIDGE AND THAT INSULIN WAS SEEN DRIPPING OUT OF THE END OF THE PLUNGER. THE FAMILY MEMBER SAID THAT THE PT FILLED A NEW CARTRIDGE FROM THE SAME LOT NUMBER AND DISCOVERED THE CARTRIDGE WAS LEAKING FROM THE BOTTOM OF THE PLUNGER. THE PT CONFIRMED THAT THE PLUNGER WAS STRAIGHT AND NOT COCKED, THE O-RINGS WERE PRESENT, THERE WERE NO VISIBLE CRACKS ALONG THE SIDES OF THE PLUNGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORP. IR 1200/1250/2020/OTP B201582

Patients

Seq Age Sex Outcome Treatment
1 15 YR