INRATIO
Report
- Report Number
- 2027969-2011-00419
- Event Type
- Injury
- Date Received
- March 1, 2011
- Date of Event
- January 31, 2011
- Report Date
- February 28, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: INRATIO: 2.1, REFERENCE: 2.0, MEAN: 2.05, CONFIDENCE LIMITS: 1.3-2.7. A 1.5 INR WAS EXCLUDED FROM COMPARISON TEST SINCE NO CORRESPONDING REFERENCE OR REPEATED INRATIO VALUE WAS PROVIDED. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT INRATIO AND REFERENCE TEST RESULT COMPARISON DID MEET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. NO FURTHER INVESTIGATION REQUIRED. AS REVIEWED ON 02/15/2011, FOURTEEN DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #243103, YIELDING A COMPLAINT RATE OF 0.016%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%), NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 1.5, LAB: NG. DATE: (B)(6) 2011, INRATIO: 2.1, LAB: 2.0. DOCTOR RAISED PATIENT'S COUMADIN DOSE ON NIGHT OF (B)(6) 2011, DUE TO INRATIO READING. PATIENT'S THERAPEUTIC RANGE: 2.0-3.0 INR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 243103 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Other |