FDA Adverse Event Malfunction Summary report: N

NOVOPEN 4 (INSULIN DELIVERY DEVICE) NA

MDR report key: 2012732 · Received March 2, 2011

Report

Report Number
9681821-2011-00009
Event Type
Malfunction
Date Received
March 2, 2011
Report Date
February 3, 2011
Manufacturer
NOVO NORDISK A/S, MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
20-986
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: THE OBSERVED PROBLEM ON THE PUSH-BUTTON IS DUE TO INCORRECT HANDLING DURING USE. COMPANY COMMENT: UPON ANALYSIS A FAULT WHICH COULD LEAD TO AN INCIDENT WAS IDENTIFIED. THE PUSH-BUTTON WAS BLOCKED AND THE PISTON ROD MOVES BACKWARDS DURING DOSAGE SELECTION. AN INTERNAL PART IN THE PEN IS BROKEN, BUT DUE TO THE PEN CONSTRUCTION THE DOSE ACCURACY IS NOT AFFECTED IF THE PEN IS USED ACCORDING TO THE INSTRUCTION. HOWEVER, IF THE CUSTOMER AFTER CHANGING THE PENFILL IS SETTING THE DOSE BEFORE RETURNING THE PISTON ROD AND NEGLECTING TO PRIME THE PEN, THE PATIENT COULD RECEIVE A TOO HIGH DOSE AND EXPERIENCE A HYPOGLYCEMIC EVENT. THE FAULT SHOULD INITIALLY BE OBVIOUS TO THE PATIENT, BECAUSE OF THE CHANGE IN INJECTION FORCE AND THE OVERALL RISK OF AN ADVERSE EVENT IS CONSIDERED MINIMAL. FINAL COMMENT FROM THE MANUFACTURER: 01/31/2010: DURING INVESTIGATION OF THE DEVICE, A FAULT WHICH COULD LEAD TO AN INCIDENT WAS IDENTIFIED. ONE CASE HAS REPORTED ADVERSE EVENTS, IN CONNECTION TO A FAULT SIMILAR TO THAT IN CASE (B)(4). IT WAS ESTIMATED THAT THE FAULT FOUND DID NOT HAVE CAUSAL RELATIONSHIP WITH ADVERSE EVENTS REPORTED IN THE CASE. THE REPORTING RATE FOR THE REPORTS, WHERE A SIMILAR FAULT AS THAT SEEN IN (B)(4) HAS BEEN FOUND IS LOW. EVALUATION SUMMARY: INVESTIGATION AND RESULTS: TECHNICAL, VISUAL, AND FUNCTIONAL EXAMINATIONS WERE PERFORMED. THE DEVICE WAS TESTED WITH A RANDOM PENFILL CARTRIDGE AND A NEW NOVOFINE NEEDLE MOUNTED. THE DOSE ACCURACY WAS MEASURED BY WEIGHING USING A RANDOM PENFILL CARTRIDGE. THE RESULT WAS WITHIN ACCEPTABLE LIMITS. THE PISTON ROD MOVES BACKWARDS DURING DOSAGE SELECTION. AN INTERNAL PART IS BROKEN, BUT DUE TO THE PEN CONSTRUCTION THE DOSE ACCURACY IS NOT AFFECTED. CONCLUSION: THE OBSERVED PROBLEM ON THE PUSH-BUTTON IS DUE TO INCORRECT HANDLING DURING USE.

Description of Event or Problem · 1

THE PISTON ROD MOVES BACKWARDS DURING DOSAGE SELECTION, AN INTERNAL PART WAS BROKEN [DEVICE MALFUNCTION]. CASE DESCRIPTION: THE INCIDENT DOES NOT REPRESENT A SERIOUS PUBLIC HEALTH THREAT. CLASS IIB. SPONTANEOUS REPORT RECEIVED FROM (B)(6) AND REPORTED BY A CONSUMER AS "THE PISTON ROD MOVES BACKWARDS DURING DOSAGE SELECTION. AN INTERNAL PART IS BROKEN, BUT DUE TO THE PEN CONSTRUCTION THE DOSE ACCURACY IS NOT AFFECTED." IT CONCERNS A PATIENT TREATED WITH NOVOPEN 4 (INSULIN DELIVERY DEVICE) (DRUG FIRST DOSE UNKNOWN, DRUG STOP DATE UNKNOWN) FOR "DEVICE THERAPY." UPON ANALYSIS OF THE RETURNED PRODUCT, A FAULT WHICH MAY LEAD TO A SEVERE MEDICAL CONSEQUENCE WAS FOUND. THIS FAULT MAY RESULT IN THE PATIENT RECEIVING TOO HIGH A DOSE AND EXPERIENCING A HYPOGLYCAEMIC EVENT. THIS CASE WAS RECLASSIFIED TO AN INCIDENT DUE TO THIS FAULT. ANALYSIS RESULT: PRODUCT NAME: NOVOPEN 4. BATCH: YUG0137. INVESTIGATION AND RESULTS: TECHNICAL, VISUAL, AND FUNCTIONAL EXAMINATIONS WERE PERFORMED. THE DEVICE WAS TESTED WITH A RANDOM PENFILL CARTRIDGE AND A NEW NOVOFINE NEEDLE MOUNTED. THE DOSE ACCURACY WAS MEASURED BY WEIGHING USING A RANDOM PENFILL CARTRIDGE. THE RESULT WAS WITHIN ACCEPTABLE LIMITS. THE PISTON ROD MOVES BACKWARDS DURING DOSAGE SELECTION. AN INTERNAL PART IS BROKEN, BUT DUE TO THE PEN CONSTRUCTION THE DOSE ACCURACY IS NOT AFFECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVOPEN 4 (INSULIN DELIVERY DEVICE) NA INSULIN DELIVERY DEVICE FMF NOVO NORDISK A/S, MEDICAL SYSTEMS NA YUG0137

Patients

Seq Age Sex Outcome Treatment
1 UNK Other