FDA Adverse Event
Malfunction
Summary report: N
PROXIMATE** SKIN STAPLER 35 WIDE
MDR report key: 2012698
·
Received March 10, 2011
Report
- Report Number
- 3005075853-2011-00953
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Report Date
- February 16, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDT
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE ANALYSIS RESULTS CONFIRMED THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND FULLY FUNCTIONAL. WHEN TESTED FOR FUNCTIONALITY, THE DEVICE FIRED AND FORMED THE REMAINING 27 STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES AND THE STAPLES WERE NOTE TO HAVE THE PROPER BOX SHAPE. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION.
Additional Manufacturer Narrative · 1
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE FIRED MALFORMED STAPLES. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE** SKIN STAPLER 35 WIDE | SKIN STAPLER | GDT | ETHICON ENDO-SURGERY, LLC. | NA | G4UC7M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |