SEE H-10
Report
- Report Number
- 2023826-2011-00142
- Event Type
- Malfunction
- Date Received
- March 9, 2011
- Date of Event
- February 15, 2011
- Report Date
- February 16, 2011
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NURSE
Narratives
SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE). (B)(4) - NO CONSEQUENCES OR IMPACT TO PATIENT. (B)(4) - LENS (IOL), TORN, SPLIT, CRACKED. EVALUATION RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND PIECES OF THE LENS OPTIC AND ONE HAPTIC TORN OFF AND MISSING. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, WERE ADDRESSED IN A CAPA OPENED IN (B)(4) 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).
THE REPORTER STATED THE SURGEON INSERTED AN AQ2015A SILICONE ASPHERIC THREE PIECE LENS AND THE LENS TORE. THE LENS WAS CUT OUT OF THE EYE TO REMOVE AND THERE WAS NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES. ANOTHER SAME MODEL LENS WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEE H-10 | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | AQ2015A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | CARTRIDGE: MODEL AQ CARTRIDGE-FP, LOT NUMBER UNK| INJECTOR: MODEL MSI-PM, LOT NUMBER UNK |