NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2011-00144
- Event Type
- Malfunction
- Date Received
- March 9, 2011
- Report Date
- February 9, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE NEOPUFF COMPONENTS WERE VISUALLY INSPECTED FOR DAMAGE. RESULTS: THE GAS INLET PORT WAS BROKEN OFF OF THE MANIFOLD. THE COLOR ON THE FRACTURED SECTION IS UNIFORM. THE END CAP ON THE REAR CASE OF THE NEOPUFF WAS DAMAGED. IN ADDITION, IT WAS OBSERVED THAT THE UPPER AND LOWER END CAPS WERE MISSING FROM THE DEVICE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090124. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE DAMAGE TO THE NEOPUFF CASE AND GAS INLET PORT IS CONSISTENT WITH IMPACT DAMAGE. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT".
(B)(4). THE COMPLAINT DEVICE HAS BEEN RECEIVED BY FISHER & PAYKEL HEALTHCARE'S REGIONAL OFFICE AND SERVICE CENTER IN (B)(4). THE COMPONENTS ARE CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPONENTS AND COMPLETION OF OUR INVESTIGATION.
A MEDICAL CENTER IN (B)(6) REPORTED THAT THE GAS OUTLET PORT ON AN RD900 NEOPUFF INFANT RESUSCITATOR WAS DAMAGED.NO PATIENT CONSEQUENCES WERE REPORTED.
A MEDICAL CENTER IN (B)(6) REPORTED THAT THE GAS OUTLET PORT ON AN RD900 NEOPUFF INFANT RESUSCITATOR WAS DAMAGED. NO PATIENT CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900AEU | 090124 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |