FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2012428 · Received March 9, 2011

Report

Report Number
2954323-2011-02145
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
February 15, 2011
Report Date
April 20, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

INVESTIGATION WAS PERFORMED ON RETURNED METER (B)(4) AND RETURNED TEST STRIP LOT NUMBER 1032124. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER MEMORY. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 22 MG/DL AND 100 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1032124

Patients

Seq Age Sex Outcome Treatment
1