FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2012389 · Received March 9, 2011

Report

Report Number
1423500-2011-02953
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
December 29, 2010
Report Date
February 17, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN: FALSE EMPTY DETECT AT INITIAL DRAIN AND AT PREVIOUS THERAPY LAST DRAIN. A REVIEW OF THE PREVIOUS SERVICE RECORD SHOWS THE DEVICE PASSED ALL REQUIRED TESTS AND CALIBRATIONS PRIOR TO ITS RELEASE. NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE ISSUES OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE EVALUATION IS IN PROCESS; HOWEVER, NOT YET COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG ON (B)(6) 2010 STARTED AT 5:28:20 WITH DRAIN VOLUME OF 3238 ML DURING CYCLE 3. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 77 YR