FDA Adverse Event Injury Summary report: N

CHOICE GUIDEWIRE

MDR report key: 2012383 · Received March 9, 2011

Report

Report Number
2134265-2011-00703
Event Type
Injury
Date Received
March 9, 2011
Date of Event
February 9, 2011
Report Date
February 9, 2011
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA
Product Code
DQX
PMA / PMN Number
K970244
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: AGE, OTHER RELEVANT HISTORY, UPN, UF/VOLUNTARY MEDWATCH#.(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR ID# 2134265-2011-00704. SAME CASE AS MDR ID# 2134265-2011-00705. IT WAS REPORTED THAT DURING AN UNSPECIFIED TREATMENT PROCEDURE, A GUIDE WIRE TIP FRACTURE OCCURRED. THE TOTALLY OCCLUDED LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A PT2 FLOPPY GUIDE WIRE WAS ADVANCED TO THE TARGET LESION. FOR BACK UP SUPPORT AN APEX BALLOON CATHETER WAS ADVANCED TO THE TARGET LOCATION. DURING ADVANCEMENT OF THE BALLOON CATHETER THE TIP OF THE PT2 FLOPPY GUIDE WIRE FRACTURED AND REMAINED INSIDE THE PATIENT. A PT2 EXTRA SUPPORT GUIDE WIRE WAS ADVANCED TO RETRIEVE THE FRACTURED TIP. HOWEVER, DURING THE RETRIEVAL ATTEMPT THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE FRACTURED "AT THE SAME AREA". THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE WAS SUCCESSFULLY REMOVED ALONG WITH THE APEX BALLOON CATHETER. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS IS STABLE.

Description of Event or Problem · 1

SAME CASE AS MDR ID# 2134265-2011-00704. SAME CASE AS MDR ID# 2134265-2011-00705. IT WAS REPORTED THAT DURING AN UNSPECIFIED TREATMENT PROCEDURE, A GUIDE WIRE TIP FRACTURE OCCURRED. THE TOTALLY OCCLUDED LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A PT2 FLOPPY GUIDE WIRE WAS ADVANCED TO THE TARGET LESION. FOR BACK UP SUPPORT AN APEX BALLOON CATHETER WAS ADVANCED TO THE TARGET LOCATION. DURING ADVANCEMENT OF THE BALLOON CATHETER THE TIP OF THE PT2 FLOPPY GUIDE WIRE FRACTURED AND REMAINED INSIDE THE PATIENT. A PT2 EXTRA SUPPORT GUIDE WIRE WAS ADVANCED TO RETRIEVE THE FRACTURED TIP. HOWEVER, DURING THE RETRIEVAL ATTEMPT THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE FRACTURED "AT THE SAME AREA". THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE WAS SUCCESSFULLY REMOVED ALONG WITH THE APEX BALLOON CATHETER. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS IS STABLE.

Description of Event or Problem · 1

SAME CASE AS MDR ID# 2134265-2011-00704. SAME CASE AS MDR ID# 2134265-2011-00705. IT WAS REPORTED THAT DURING AN UNSPECIFIED TREATMENT PROCEDURE, A GUIDE WIRE TIP FRACTURE OCCURRED. THE TOTALLY OCCLUDED LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A PT2 FLOPPY GUIDE WIRE WAS ADVANCED TO THE TARGET LESION. FOR BACK UP SUPPORT AN APEX BALLOON CATHETER WAS ADVANCED TO THE TARGET LOCATION. DURING ADVANCEMENT OF THE BALLOON CATHETER THE TIP OF THE PT2 FLOPPY GUIDE WIRE FRACTURED AND REMAINED INSIDE THE PATIENT. A PT2 EXTRA SUPPORT GUIDE WIRE WAS ADVANCED TO RETRIEVE THE FRACTURED TIP. HOWEVER, DURING THE RETRIEVAL ATTEMPT THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE FRACTURED "AT THE SAME AREA". THE TIP OF THE PT2 EXTRA SUPPORT GUIDE WIRE WAS SUCCESSFULLY REMOVED ALONG WITH THE APEX BALLOON CATHETER. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHOICE GUIDEWIRE WIRE, GUIDE, CATHETER DQX BOSTON SCIENTIFIC - COSTA RICA H74912160011

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention