FDA Adverse Event Malfunction Summary report: N

VISIAN ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 2012368 · Received March 9, 2011

Report

Report Number
2023826-2011-00203
Event Type
Malfunction
Date Received
March 9, 2011
Report Date
February 10, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P030016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION METHOD: DEVICE HISTORY RECORD REVIEW. RESULTS: A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED AND NOTHING WAS FOUND IN THE MANUFACTURING PROCESS OF THIS LENS THAT WAS THE ROOT CAUSE OF THE COMPLAINT. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, DEVICE HISTORY RECORD REVIEW AND THE EVALUATION OF THE RETURNED PRODUCT, THE MOST PROBABLE CAUSE OF THE EVENT WAS MANIPULATION RELATED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4) - LENS DAMAGED IN DELIVERY SYSTEM. EVALUATION METHOD: LENS WORK ORDER SEARCH. RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A PIECE OF ONE HAPTIC TORN OFF AND MISSING. THE LENS WAS RETURNED IN LIQUID. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: NO CONCLUSION CAN BE DRAWN. BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON ATTEMPTED TO INSERT A 12.6MM MICL 12.6 IMPLANTABLE COLLAMER LENS, BUT THE LENS BROKE UPON INSERTION INTO THE INJECTOR. THE BACK-UP LENS WAS IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VISIAN ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY MICL 12.6

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR: MODEL AND LOT NUMBER UNK| CARTRIDGE: MODEL AND LOT NUMBER UNK