FDA Adverse Event Malfunction Summary report: N

HEARTMATE II SYSTEM CONTROLLER

MDR report key: 2012168 · Received February 1, 2011

Report

Report Number
2916596-2011-00043
Event Type
Malfunction
Date Received
February 1, 2011
Date of Event
October 14, 2010
Report Date
January 5, 2011
Manufacturer
THORATEC CORP.
Product Code
DSQ
Removal / Correction Number
29165969/2/10001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM CONTROLLER WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE REPORTED EVENT WAS CONFIRMED. FUNCTIONAL TESTING OF THE RETURNED SYSTEM CONTROLLER REVEALED THAT WHEN THE BLACK POWER LEAD WAS MANIPULATED AT THE CONNECTOR END DURING TESTING, THE PUMP STOPPED, THE POWER MODULE SOUNDED A CONTINUOUS AUDIO ALARM AND THE RED AND YELLOW BATTERY LEDS ON THE SYSTEM CONTROLLER WERE ILLUMINATED. THE SYSTEM CONTROLLER WAS DISCONNECTED FROM THE TEST SYSTEM AND UPON INSPECTION OF THE INNER CONDUCTORS AT THE CONNECTOR END OF THE BLACK POWER LEAD, THE BROWN CONDUCTOR THAT MONITORS THE BATTERY VOLTAGE WAS FOUND TO BE SEVERED. A REVIEW OF DEVICE HISTORY RECORDS SHOWED NO DEVIATIONS FROM MANUFACTURING OR QA SPECIFICATIONS. THIS SITUATION IS BEING ADDRESSED THROUGH THE MANUFACTURER'S CORRECTIVE/PREVENTATIVE ACTION SYSTEM AND AN URGENT MEDICAL DEVICE CORRECTION NOTICE WAS SENT TO CUSTOMERS. NO FURTHER INFORMATION IS AVAILABLE. THE MANUFACTURER IS CLOSING ITS FILE ON THIS EVENT.

Description of Event or Problem · 1

THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). THE VAD COORDINATOR REPORTED THAT THE PATIENT WAS EXPERIENCING LOW VOLTAGE AND POWER CABLE DISCONNECT ALARMS. THE SUSPECT SYSTEM CONTROLLER WAS EXCHANGED PER THE MANUFACTURER'S INSTRUCTIONS FOR USE AND THE PATIENT REMAINS ONGOING ON LVAD SUPPORT WITH NO FURTHER ISSUE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II SYSTEM CONTROLLER LVAD SYSTEM CONTROLLER DSQ THORATEC CORP. 103696 NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other